DSO152 Digital Oscilloscope The
DSO152 is a digital oscilloscope with a 2.8-inch 320*240 resolution
high-definition LCD screen. It features a built-in 1000mAh high-quality
lithium battery, which can be used continuously for about 4 hours after
being fully charged.Key Features:
- Sampling rate: 2.5MS/s
- Bandwidth: 200K
- Vertical sensitivity: 10mV/Div-20V/Div (Progress according to the 1-2-5 way)
- Time base range: 10s/Div-50s/Div (Progress according to the 1-2-5 way)
- Voltage range: X1:40V (Vpp:80V), X10: 400V (Vpp :800V)
- Trigger method: Auto/Normal/Single
- Coupling method: AC/DC
- Display: 2.8 inches/PPI:320*240
- USB charging: 5V/1A
- Lithium battery capacity: 1000mAh
- Square wave calibration:
- Frequency: 1K
- Duty cycle: 50%
- Size: 99×68.3×19.5mm
- Weight: 100g
Product Usage Instructions
Panel Introduction The DSO152 digital oscilloscope has the following panel features:
- Signal input probe (MXC) interface
- Square wave calibration
- Trackwheel buttons (Left, OK, Right)
- Reset hole
- Charging indicator light (red light during charging, green light when fully charged)
- TYPE-C charging interface (5V/1A)
- 2.8-inch LCD screen
- AUTO button
- MODE button
- Up button
- Down button
- RUN button
- Switch button
Key Function The trackwheel buttons on the DSO152 digital oscilloscope have the following functions:
Button |
Operation |
Function |
Short press |
|
Control parameters function selection |
|
Parameter addition adjustment |
Long press |
|
Exit auto calibration (Auto calibration page) |
|
Parameter subtraction adjustment |
Short press |
|
Short press |
Long press |
|
Run/pause waveforms (other pages) |
Short press |
|
Enter auto calibration (Auto calibration page) |
Short press |
|
Show/close detailed parameters |
Screen Instructions The DSO152 digital oscilloscope provides the following screen instructions:
- Trigger mode indicator icon: Auto (automatic trigger), Single (single trigger), Normal (normal trigger)
- Baseline indicator icon: Indicates the current position is OV voltage
- Vertical sensitivity: Represents the voltage by a large grid in the vertical direction
- 1X/10X mode indicator icon: Must be consistent with the 1X/10X switch the setting on the probe handle
Notice to users
- This
manual introduces the usage method, precautions, and related matters of
the product in detail. Before using the product, please read the manual
carefully in order to give full play to the best performance of the
product.
- Do not use the device in a flammable or explosive environment.
- The
used batteries are replaced by the device and the discarded device
cannot be disposed of together with domestic waste. Please dispose of
them according to relevant national or local laws and regulations.
- When
there is any quality problem with the device or you have any questions
about its use of it, please contact the online customer service or the
manufacturer of “FNIRSI”, and we will solve it for you in the first time.
Product introduction
“DSO152” is a highly practical and cost-effective handheld oscilloscope launched
by our company, which is aimed at the maintenance industry and the
research education industry. The oscilloscope has a real-time sampling
rate of 2.5MS/s, a bandwidth of 200KHz, and complete trigger functions
(single, normal, and automatic). It can be used freely for both periodic
analog signals and non-periodic digital signals and can measure voltages up to
±400V. Equipped with an efficient one-key AUTO, it can
display the measured waveform without cumbersome adjustments. Equipped
with a 2.8-inch 320*240 resolution high-definition LCD screen. Built-in
1000mAh high-quality lithium battery, can be used continuously for about
4 hours after fully charged.
Key Function
Button |
Operation |
Function |
|
Short press |
Control parameters function selection |
|
Short press |
Exit auto calibration (Auto calibration page) |
Long press |
Enter the automatic calibration page |
|
Short press |
Control parameters function selection |
AUTO |
Short press |
Automatic adjustment (frequency below 45Hz cannot be calibrated correctly) |
MODE |
Short press |
AUTO/Single/Normal switching |
Long press |
Rising edge/falling edge switching |
|
Short press |
Parameter addition adjustment |
|
Short press |
Parameter subtraction adjustment |
RUN |
Short press |
Run/pause waveforms (other pages) Enter auto calibration (Auto calibration page) |
Long press |
Show/close detailed parameters |
|
Short press |
OFF |
Long press |
On |
Parameters
Sampling rate |
2.5MS/s |
Bandwidth |
200K |
Vertical sensitivity |
10mV/Div-20V/Div(Progress according to the 1-2-5 way) |
Time base range |
10μs/Div-50s/Div(Progress according to the 1-2-5 way) |
Voltage range |
X1:士40V (Vpp:80V) |
X10: 士400V (Vpp :800V) |
Trigger method |
Auto/Normal/Single |
Coupling method |
AC/DC |
Display |
2.8 inches/PPI:320*240 |
USB charging |
5V/1A |
Lithium battery capacity |
1000mAh |
Square wave calibration |
Frequency: 1K Duty cycle: 50% |
Size |
99×68.3×19.5mm |
Weight |
100g |
Screen instructions
- Trigger mode indicator icon, Auto means automatic trigger, Single means single trigger, Normal means normal trigger.
- Baseline indicator icon, this icon indicates the current position is OV voltage.
- Vertical sensitivity, which means the voltage represented by a large grid in the vertical direction.
- 1X/10X
mode indicator icon, this must be consistent with the 1X/10X switch
setting on the probe handle if the probe is in 1X mode, then the
oscilloscope should also be set to 1X mode, 1X measures ±40V voltage,
10X measures ±400V voltage.
- Horizontal time base, indicating the length of time represented by a large grid in the horizontal direction.
- Input coupling indicator icon, AC means AC coupling and DC means DC coupling.
- Pause running indicator icon, RUN means running, STOP means pause.
- Trigger voltage indicator icon
- Trigger edge indicator icon
Precautions
- After receiving the device, please use it after fully charged.
- When
using the oscilloscope, pay attention to the selection of the gear, the
gear of the oscilloscope should be consistent with the gear of the
probe.
- When measuring high voltage, it is forbidden to touch any metal part of the oscilloscope to avoid the risk of electric shock.
- Try not to perform a high-voltage test while charging.
- When calibrating, need to unplug the BNC probe, or short the positive and negative poles of the probe.
|